Haftungsausschluss: Diese Website bündelt öffentlich verfügbare Daten aus offiziellen Quellen wie FDA, ClinicalTrials.gov, PubMed und SEC EDGAR nur zur allgemeinen Information. Sie stellt keine medizinische Beratung, Diagnose, Behandlungsempfehlung oder Anlageberatung dar.

PubMed

Research: User Interface Software Errors in Medical Devices: Study of U.S. Recall Data.

PMID: 31162965 · 2019

Y, i, , Z, h, a, n, g, ;, , P, a, o, l, o, , M, a, s, c, i, ;, , P, a, u, l, , J, o, n, e, s, ;, , H, a, r, o, l, d, , T, h, i, m, b, l, e, b, y

ZeitschriftBiomedical instrumentation & technology
Jahr2019
PMID31162965

Zusammenfassung

The current work assessed U.S. medical device recalls during 2012-15, with the goal of under-standing the impact and nature of user interface (UI) software errors in medical devices. Based on information from the Food and Drug Administration's public and internal recall databases, 423 (~140/year) medical device recalls were identified as resulting from UI software errors, which accounted for nearly one-half of recalls caused by software errors during the same period. A total of 499 UI software e

Offizielle Quelle

Auf PubMed ansehen →

Daten aus PubMed / NCBI. Den Volltext und die aktuellsten Informationen finden Sie stets im offiziellen Eintrag.

↑