Evaluation of Ophthalmic Surgical Instrument Sterility Using Short-Cycle Sterilization for Sequential Same-Day Use.

PMID: 29602568 ·

D, a, v, i, d, , F, , C, h, a, n, g, ;, , N, i, k, k, i, , H, u, r, l, e, y, ;, , N, i, c, k, , M, a, m, a, l, i, s, ;, , J, e, f, f, r, e, y, , W, h, i, t, m, a, n

RevistaOphthalmology
Año
PMID29602568

Resumen

The common practice of short-cycle sterilization for ophthalmic surgical instrumentation has come under increased regulatory scrutiny. This study was undertaken to evaluate the efficacy of short-cycle sterilization processing for consecutive same-day cataract procedures. Testing of specific sterilization processing methods by an independent medical device validation testing laboratory. Phaco handpieces from 3 separate manufacturers were tested along with appropriate biologic indicators and contr