Aviso: Este site agrega dados públicos oficiais da FDA, ClinicalTrials.gov, PubMed e SEC EDGAR apenas para referência geral. Não constitui aconselhamento médico, diagnóstico, recomendação de tratamento ou aconselhamento de investimento.

Ensaio clínico

Accuracy of a Cuffless Photoplethysmography (PPG) Chest-Patch Monitor for 24-hour Ambulatory Blood Pressure Monitoring (ABPM)

NCT: NCT07240831 · COMPLETED

ID NCTNCT07240831
EstadoCOMPLETED
Data de início2025-12-15
Conclusão2026-09-02
SponsorBiobeat Technologies Ltd. (INDUSTRY)
First posted
Last updated

Resumo breve

This study aims to validate the accuracy and reliability of blood pressure (BP) estimates obtained over 24 hours using a PPG-based chest-patch device compared to the gold standard ambulatory blood pressure monitoring (ABPM) method using an upper arm cuff-based oscillometric BP device, in both hypertensive and normotensive individuals referred by their provider to undergo a 24-hours ABPM for clinical indication. The Awake/Asleep test, which is the primary test recommended for automated wearable cuffless BP devices that are cuff-calibrated (based on the 2023 European Society of Hypertension (ESH) recommendations for the validation of cuffless blood pressure measuring devices), will be conducted in this study. The secondary aim of the study is to assess the feasibility and convenience of the PPG-based device.

Fonte oficial

Ver no ClinicalTrials.gov →

Dados provenientes da API do ClinicalTrials.gov. Para consultar o estado mais recente, veja o registo oficial.

↑