Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR, NMPA) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

PubMed

CORE-MD clinical risk score for regulatory evaluation of artificial intelligence-based medical device software.

PMID: 39915308 · 2025

JournalNPJ digital medicine
Year2025
PMID39915308

Abstract

The European CORE-MD consortium (Coordinating Research and Evidence for Medical Devices) proposes a score for medical devices incorporating artificial intelligence or machine learning algorithms. Its domains are summarised as valid clinical association, technical performance, and clinical performance. High scores indicate that extensive clinical investigations should be undertaken before regulatory approval, whereas lower scores indicate devices for which less pre-market clinical evaluation may

Official Source

View on PubMed →

Data sourced from PubMed / NCBI. For the full text and most current information, always refer to the official record.