Sedecal SP-HF 4.0 Portable X-Ray System; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) Spain
FDA Recall Z-0001-04 · Posted 2003-10-29
Advertisement
Recalling FirmSedecal USA, Inc
Recall StatusTerminated
Product CodeIZL
Date Initiated2003-10-27
Date Terminated2004-03-24
Product Quantity1 unit
Reason for Recall
A minimum source-skin-distance of less than 30 cm and not identifying the tube manufacturer on the tube housing label resulted in the SP-HF-4.0 Portable Systems not complying with the U.S. Federal performance standard.
Firm Action
An 'Important Safety Notice' dated 10/27/03 was sent to the one direct account on the same date, informing them of the non-compliances with the performance standard and requesting them to call Sedecal at 800-920-9525 to arrange for the modification of the x-ray unit at no cost to the user.
Root Cause
Other
Distribution Pattern
Vermont
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.