CDI H/S Cuvette The intended use for the CDI H IS Cuvette is as follows: The CDI H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
FDA Recall Z-0001-2017 · Posted None
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeDRY
Date Initiated2015-12-11
Date Terminated2016-10-26
Product Quantity140,000 units
Reason for Recall
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
Firm Action
Terumo (Planned Action) plans to execute the following steps as part of the correction activity:
Notify all customers of the potential that a cuvette may disengage from the circuit (probe) leading to a "disconnect" message on the monitor. This notification will be performed via a Customer Notification Letter.
For further questions, please call (410) 392-7277.
Root Cause
Other
Distribution Pattern
Worldwide Distribution - USA (nationwide) Distribution.