Terminated

Revolution XR/d system control room PC model 5117866-2 with Software collector 5115434

FDA Recall Z-0002-2009 · Posted 2009-09-24

Recall StatusTerminated
Product CodeKPR
Date Initiated2008-02-14
Date Terminated2012-01-07
Product Quantity129

Reason for Recall

GE Healthcare has recently become aware of a potential risk of observing a blank image after exposure on Revolution XR/d systems equipped with HP 8200 PC user consoles that may impact patient safety. It was reported that images acquired with off-center collimation for extremities with the position selected as Digits to front or Digits to back are displayed as blank. Investigation into th

Firm Action

Consignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution -- AL, AZ, CA, CO, FL, GA, HI, IL , IN, KY, LA, MD, MA, MI,MN,MS, MO, MT, NE, NH, NJ, NM, NY, NC, OH, OK OR, PA, SC, TN, TX, VA, WA, and WV.