Merit Inflation Syringe Kit. Catalog Number KOS-02857

FDA Recall Z-0002-2017 · Posted None

Recall StatusTerminated
Product CodeMAV
Date Initiated2016-08-25
Date Terminated2016-12-01
Product Quantity1,297 total units

Reason for Recall

Merit Medical Systems, Inc. announces a voluntary field action for Merit Inflation Syring Kit due to unit labels missing required device labeling information.

Firm Action

Merit Medical Inc. notified all affected customers on August 25, 2016, via verbal and written notification to return the affected units. The returned units will be re-labeled. For further questions, please call (801) 316-3690.

Root Cause

Labeling design

Distribution Pattern

One consignee located in CA. No Canadian, other foreign or VA/govt/military consignees.