Revolution XR/d systems with control room PC model 5117866-2 with Software collector 5194381. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188. Intended for use in generating Tomographic images of human anatomy.

FDA Recall Z-0003-2009 · Posted 2009-09-22

Recall StatusTerminated
Product CodeKPR
Date Initiated2008-02-14
Date Terminated2011-12-18
Product Quantity83 (63 USA, 20 OUS)

Reason for Recall

GE Healthcare has recently become aware of an error in the generator software on the Revolution XR/d configured with HP 8200 PC consoles that reports a mAs reading that is higher than is actually exposed to patients.

Firm Action

Consignees were sent a GE Healthcare "Product Safety Notification" letter dated March 7, 2008. The letter was addressed to Hospital Administrators, Mangers of Radiology, and Radiologists. The letter listed the Safety Issue, Affected Product, Safety Instructions, and Contact Information.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA, including states of AL, AZ, CA, CO, FL, ID, IL, IN, LA, MA, MI, MO, NY, NC, OH, OR, PA, SC, SD, TN, TX, WA, and WI and countries of Canada, China, Japan, Korea, and Taiwan.