Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BVI 9600 Bladder Volume Instrument, Aorta Scan Mode The BladderScan BVI 9600 with AortaScan Mode is a user-selectable, dual-function ultrasound device that projects ultrasound energy either into the lower abdomen to obtain an image of the bladder for measuring bladder volume, or into the mid-abdomen to obtain an image of the abdominal aorta for aortic diameter measurements.

FDA Recall Z-0003-2011 · Posted 2010-10-04

Recalling FirmVerathon, Inc.
Recall StatusTerminated
Product CodeIYO
Date Initiated2010-08-23
Date Terminated2011-09-08
Product Quantity351 units

Reason for Recall

The BladderScan BVI 9600 with AortaScan devices may experience the loss of a factory installed software configuration file which results in a loss of the aorta measurement function.

Firm Action

Verathon sent an URGENT Medical Device Recall Notification letter dated August 12, 2010 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Verathon provided an addendum to the User Manual with instructions on how to reset the instrument. Customers were advised to: " Add the enclosed addendum to their current Users Manual. " Update the application software of the device to v2.7.0 or higher. If the customers had questions or needed support for the software update process, they were to contact Verathon Medical Customer Care at (425) 867-1348 or (800) 331-2313 or [email protected].

Root Cause

Software design

Distribution Pattern

Worldwide Distribution - USA and the countries of Australia, Canada, China, Colombia, European Union, United Kingdom, Saudi Arabia, and Singapore.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑