Fortify VR - Model Numbers: CD1231-40 and CD1231-40Q; Fortify ST VR, Model Numbers: CD1241-40 and CD1241-40Q; Fortify Assura VR, Model Numbers: CD1257-40, CD1257-40Q, CD1357-40C, CD1357-40Q; Fortify Assura ST VR, Model Numbers: CD1263-40, CD1263-40Q, CD1363-40C, CD1363-40Q; Fortify DR, Model Numbers: CD2231-40 and CD2231-40Q; Fortify ST DR, Model Numbers: CD2241-40 and CD2241-40Q; Fortify Assura DR, Model Numbers: CD2257-40, CD2257-40Q, CD2357-40C, CD2357-40Q; Fortify Assura ST DR, Model Numbers: CD2263-40, CD2263-40Q, CD2363-40C, CD2363-40Q Product Usage Fortify and Fortify Assura implantable cardioverter defibrillators (ICDS) are intended to provided ventricular anti-tachycardia pacing and ventricular defibrillation for automated treatment of life threatening ventricular arrhythmias. The devices may also be programmed to provide single or dual chamber bradycardia pacing
FDA Recall Z-0003-2018 · Posted 2017-10-10
Advertisement
Recalling FirmSt Jude Medical Inc.
Recall StatusTerminated
Product CodeLWS
Date Initiated2017-08-28
Date Terminated2024-01-20
Product Quantity90,563 units
Reason for Recall
The device may exhibit premature battery depletion.
Firm Action
Abbott sent a Field notification letter dated August 28, 2017 to customers of the availability of the new Battery Performance Alert.
The letter identified the affected product, problem and actions to be taken. The recalling firm posted a public statement on their website in the Abbott Media Room. The link to the statement is as follows: http://abbott.mediaroom.com/2017-08-29-Abbott-issues-new-updates-for-implanted-cardiac-devices.
For questions contact your local Sales Representative or Abbott Technical Services at 1-800-722-3774.
Root Cause
Device Design
Distribution Pattern
US Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.