Hard shell carry case for the LIFEPAK 500 AED.

FDA Recall Z-0004-06 · Posted 2005-10-04

Recall StatusTerminated
Product CodeMKJ
Date Initiated2005-07-12
Date Terminated2006-06-26
Product Quantity4608 cases

Reason for Recall

Some hard shell carry case for LIFEPAK 500 and LIFEPAK CR Plus automated external defibrillators may have blocked pressure vent which may make it difficult to open the case after a sudden decrease in altitude (such as in aircraft landng or mountain descent).

Firm Action

On 7/12/05 the firm issued a letter to customers instructing the customers to inspect the cases for blocked vent. If the customer noted that the vent was blocked, to contact the firm for a replacement case. The firm provided response cards for the customers to return stating they had inspected their units.

Root Cause

Process control

Distribution Pattern

Cases were distributed nationwide and internationally.