Thoratec brand TLC-II Portable Ventricular Assist Device (VAD) Driver; Catalog No. 20010-2085-000 Distributed and Manufactured by Thoratec Corporation, 6035 Stoneridge Drive, Pleasanton, CA 94588

FDA Recall Z-0004-2007 · Posted 2006-10-03

Recalling FirmThoratec Corp
Recall StatusTerminated
Product CodeDSQ
Date Initiated2006-08-25
Date Terminated2007-11-27
Product Quantity114 devices (57 valves/2 per unit)

Reason for Recall

Sticky Valve Disk- The valve disk has an increased tendency to stick during operation causing noise and pressure/vacuum alarms.

Firm Action

On 8/26/06, the firm issued letters to all its direct consignees, informing them of the affected product and providing 'IMPORTANT PRODUCT SAFETY BULLETIN'.

Root Cause

Other

Distribution Pattern

Worldwide distribution-USA including states of AL, CA, CT, FL, IL, IN, KY, LA, MA, ME, MI, MN, NC, NY, OH, PA, TN, TX, VA, WA, and WI., with 19 Foreign consignees Australia, Belgium, Canada*, Czech Republic, France, Germany, Ireland, Netherlands, Portugal, Switzerland, United Kingdom. (*Hospital Royal Victoria, Queen Elizabeth II/Capital Health, St. Paul's Hospital, Vancouver, Canada.)--