Toshiba Infinix fluoroscopy DFP-8000D Digital Radiography System with version 3.40 or higher software. Controller for Infinix interventional angiography systems.
FDA Recall Z-0005-2009 · Posted 2009-09-28
Recalling FirmToshiba American Medical Systems Inc
Recall StatusTerminated
Product CodeMQB
Date Initiated2008-09-29
Date Terminated2012-10-04
Product Quantity129
Reason for Recall
Audible signal is not available at all times when using the HLC fluoroscopy mode. Special means of activation are not designed in for selection of High Level Controls, as required by regulation.
Firm Action
Toshiba issued a letter entitled Important Electronic Product Radiation Warning, September 30, 2008, explaining the issue. The customers were requested to reply to the letter via fax-back form, and a Toshiba service representative would install the corrective action.
Root Cause
Software design
Distribution Pattern
Nationwide