GE OEC 9900 Elite GSP Fluoroscopes To produce mobile fluoroscopic images of human anatomy.
FDA Recall Z-0005-2010 · Posted 2010-01-20
Advertisement
Recalling FirmGe Healthcare
Recall StatusTerminated
Product CodeJAA
Date Initiated2009-08-12
Date Terminated2012-04-23
Product Quantity1411 units total
Reason for Recall
A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.
Firm Action
An Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide and Canadian Distribution