Terminated

The product is labeled in part "***Surgical Laser Technologies, Inc. Montgomeryville, PA***SLT Fiber Delivery System***Sterile***Non-Pyrogenic***. The product is packaged in a flexible pouch and includes the following laser surgery accessories - SLT-SSRH 8, Catalog/Product Number 0041-3082; SLT-SSRH 11, Catalog/Product Number 0041-3112; SLT-SSRH 4-SMA, Catalog/Product Number 0042-0075; SLT-SSRH 5-SMA, Catalog/Product Number 0042-0076; SLT-SSRH 6-SMA, Catalog/Product Number 0042-0077; SLT-SSRH 7-SMA, Catalog/Product Number 0042-0078; SLT-SSRH 8-SMA, Catalog/Product Number 0042-0079; SLT-SSRH 9-SMA, Catalog/Product Number 0041-4931; and SLT-SSRH 10-SMA, Catalog/Product Number 0042-0081. Indicated for the incision, excision, coagulations, and vaporization of soft tissue.

FDA Recall Z-0005-2014 · Posted 2013-10-01

Recalling FirmPhotoMedex, Inc.
Recall StatusTerminated
Product CodeGEX
Date Initiated2013-01-04
Date Terminated2014-04-16
Product Quantity549

Reason for Recall

Firm became aware that some products within the specified lot may not have been sealed appropriately, thus compromising the sterile barrier.

Firm Action

The firm initiated their recall of these products beginning on January 4, 2013 by sending notification letters to their consignees via electronic mail and/or facsimile. Customers were asked to return the affected product for a replacement.

Root Cause

Process change control

Distribution Pattern

Worldwide Distribution-USA including the states of AR, KS and OH, and the countries of Germany, South Africa and the United Kingdom.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑