Z-800 Infusion System, Model Numbers Z-800, Z-800F, Z-800W, Z-800WF; Software Version: Z-800 6.1.01 and 6-1.07z; Z-800F 4.1.02 and 4.1.08z; Z-800W 3.1.32 and 3.1.64z, Z-800WF 3.1.32 and 3.1.64z
FDA Recall Z-0005-2025 · Posted 2024-10-09
Advertisement
Recalling FirmZyno Medical LLC
Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2024-09-13
Product Quantity34,994 units (1819 units still need correction)
Reason for Recall
There is a defect in the air-in-line software algorithm.
Firm Action
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated 9/13/24 was sent to customers.
Mitigation
Zyno Medical LLC is mitigating this software issue by conducting a voluntary medical device correction to remove the faulty software and correctly install software that is free of this defect.
Recommended Actions
1. Actions to Address the Issue During Use
Note that only the devices listed in this letter are affected. No other products are involved in this action.
Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below.
2. Actions to Correct the Issue
Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF Infusion Pump Air-in-Line Software Issue Response Verification Form. This will include checking your inventory for the affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form.
A representative from Zyno Medical s partner company, Intuvie LLC, will be in contact with you to coordinate instructions on exchanging devices to remediate this action.
Pass this notice to the appropriate personnel who need to be informed if the identified devices have been transferred to another location.
If you have any questions or would like assistance, please contact your local Business Development Manager.
Contact Information: [email protected]
***Update 1/3/2025***Additional consignees were sent URGENT: CLASS 1 RECALL letters due to additional affected units identified. Recommended Actions
1. Actions to Address the Issue During Use
Note that only the devices listed in this letter are affected. No other products are involved in this action.
Discontinue use of the affected device and proceed to the instructions for return of the device in Recommended Action below.
2. Actions to Correct the Issue
Complete the attached Zyno Medical Z-800, Z-800F, Z-800W, and Z-800WF I
Root Cause
Process change control
Distribution Pattern
US Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.