Rultract Surgical Retractor, 4100 Series, all system configurations, reusable.
FDA Recall Z-0008-04 · Posted 2003-10-15
Advertisement
Recalling FirmRultract Inc
Recall StatusTerminated
Product CodeDWS
Date Initiated2003-09-18
Date Terminated2007-03-23
Product Quantity2,718 units
Reason for Recall
Component spring or nut may loosen, detach, and fall into chest of patient during surgical procedure.
Firm Action
The firm notified their affected consignees by letter on 9/18/2003.
Root Cause
Other
Distribution Pattern
The device was distributed to end-users located nationwide.