Synchron LX Systems Version 4.5 Operating Software

FDA Recall Z-0009-06 · Posted 2005-10-06

Recalling FirmBeckman Coulter Inc
Recall StatusTerminated
Product CodeJJE
Date Initiated2005-09-06
Date Terminated2012-01-26
Product Quantity868

Reason for Recall

Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results may be reduced from above the upper limit of the laboratory's reference interval to below the limit.

Firm Action

Customers were notified by letter dated September 6, 2005 with work around instruction provided in the letter.

Root Cause

Software design

Distribution Pattern

Nationwide and Canada