EPIK KNEE SYSTEM Angled Insert Impactor LM/RL; Catalog #801-03-051 Rev. A; non-sterile; knee prosthesis instrument. Distributed by Encore Medical, L.P., Austin, TX 78758.
FDA Recall Z-0009-2008 · Posted 2007-10-10
Advertisement
Recalling FirmEncore Medical, Lp
Recall StatusTerminated
Product CodeHRY
Date Initiated2007-08-09
Date Terminated2008-03-10
Product Quantity5 units.
Reason for Recall
Angled Insert Impactor may scratch Tibial Insert on the articulating surface, causing increased wear at the damaged site.
Firm Action
Firm notified consignees regarding recall via letter on 08/09/07. Consignees informed that Encore customer service representatives would be contacting them to exchange recalled units with replacements.
Root Cause
Device Design
Distribution Pattern
Products distributed to direct accounts in CA, FL and MD.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.