Power cord manufactured by ElectriCord Manufacturing Company , P/N 316579, used with DINAMAP Patient Monitors as Follows: DINAMAP PRO Series, DINAMAP PRO 1000 Monitor under both the Critikon and GE Medical Systems. The company name EliectriCord is on the power cord plug.

FDA Recall Z-0011-04 · Posted 2003-10-15

Recall StatusTerminated
Product CodeMWI
Date Initiated2003-04-30
Date Terminated2012-05-03
Product Quantity16,339 Power Cords

Reason for Recall

Reports of broken plug ground pins on AC power cords for patient monitors.

Firm Action

The recalling firm mailed a Recall Notification Letter with response form and corretive power cords beginning on April 30, 2003 until August 14, 2003 as corrective cords became available. Consignees were instructed on how to determine if they have the recalled cords and to destroy any found by cutting the plug from the wire. Also corrective AC power cords were provided to each consignee.

Root Cause

Component design/selection

Distribution Pattern

The recalled AC power cords were distributed to approximately 2600 Domestic hospital accounts nationwide and 180 international accounts.