ellex I.Science iTRACK 250A, Canaloplasty Microcatheter Kit, for single use, REF iT-250A, Rx. The firm name on the label is Ellex iScience, Inc., Fremont, CA. Used in ophthalmic surgery.

FDA Recall Z-0011-2019 · Posted None

Recalling FirmEllex iScience, Inc.
Recall StatusTerminated
Product CodeHMX
Date Initiated2018-01-11
Date Terminated2018-10-02
Product Quantity90 units

Reason for Recall

Sterility failure found in one of the units in the lot.

Firm Action

The recalling firm called their consignees on 1/11/2018 instructing them to quarantine unused devices and monitor the patients operated with the devices in this lot. The firm issued a letter dated 1/16/2018 via FedEx Express on 1/17/2018 informing them a potential issue had been identified and to segregate the product immediately.

Root Cause

Process control

Distribution Pattern

Distribution US nationwide and Switzerland.