Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.

FDA Recall Z-0012-04 · Posted 2003-10-15

Recall StatusTerminated
Date Initiated2003-05-08
Date Terminated2003-12-09
Product Quantity478

Reason for Recall

Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.

Firm Action

Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.

Root Cause

Other

Distribution Pattern

United States