PRISM MEDICAL A-625, MAX. LOAD 625 lbs. 283 Kgs., Patient lift, non-A/C powered
FDA Recall Z-0013-2018 · Posted 2017-09-15
Advertisement
Recalling FirmHandicare Usa Inc
Recall StatusTerminated
Product CodeFSA
Date Initiated2017-09-08
Date Terminated2018-12-17
Product Quantity8 units
Reason for Recall
During internal cycle testing of A-Series lifts Handicare has found that the A-Series lift strap for lifts manufactured between April 25, 2017 and June 12, 2017 may wear prematurely at max load (625 lbs). We have identified the root cause as a supplied component that was not to specification. If the strap wears during use there is a potential risk for the person being supported by the lift to drop.
Firm Action
The firm initiated their recall by email on 09/08/2017. The consignee was warned that the lift straps may wear prematurely and they would be contacted by a representative and the units would be replaced.
Root Cause
Component design/selection
Distribution Pattern
Us and Canada
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.