Brilliance(a) CT 6 slice, Model #4535 670 73191. Brilliance(a) CT 10 Slice, Model 4535 670 73181. Brilliance 16 Slice, Model #4535 670 78851. Brilliance CT 16 Power, Model #4535 670 23331. Brilliance CT 40 Slice, Model #4550 110 04011. Brilliance CT Big Bore, Model #4535 670 88051. Computed tomography systems.

FDA Recall Z-0014-2007 · Posted 2007-09-26

Recall StatusTerminated
Product CodeKPS
Date Initiated2006-11-06
Date Terminated2010-10-05
Product Quantity20

Reason for Recall

A potential anomaly was identified on the CT subsytem, which could result in the radiation dose being different than the calculated dose reported to the operator.

Firm Action

The customers were sent Public Safety Notification November 3, 2006, indicating there was a problem with the software. However, since their unit was recently upgraded the problem did not impact their unit.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Australia, Canada, China, England, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Netherlands, Poland, Puerto Rico, Spain, and Taiwan.