Terminated

Brain Heart Infusion Agar labeled BHIA + Vancomycin 6 MCG/ML For Screening for Vancomycin Resistant Enterococci PML microbiologicals Wilsonville, Oregon 97070 USA Catalog #P1174

FDA Recall Z-0014-2008 · Posted 2007-10-10

Recall StatusTerminated
Product CodeJSO
Date Initiated2007-04-23
Date Terminated2011-06-07
Product Quantity640 total plates, 320 per lot (lot 229957-1 was dist only in Canada; lot 229958-1 was dist in the US and Canada)

Reason for Recall

Plates manufactured without vancomycin. Possibility for false resistant results.

Firm Action

On 4/23/07 the firm began contacting customers via fax and telephone.

Root Cause

Other

Distribution Pattern

lot #229957-1: 120 plates were distributed to 9 Canadian consignees; 200 plates were shipped, quarantined, and destroyed at the Canadian distribution center. lot #229958-1: 110 plates were distributed to 9 US consignees; 210 plates were shipped, quarantined, and destroyed at the Canadian distribution center.