Terminated

Synthes Midfoot Fusion Bolt 6.5 mm. It is indicated for fracture, fixation, osteotomies, nonunions, and fusions of large bones in the foot and ankle.

FDA Recall Z-0014-2014 · Posted 2013-10-17

Recalling FirmSynthes USA HQ, Inc.
Recall StatusTerminated
Product CodeHWC
Date Initiated2013-03-08
Date Terminated2015-08-12
Product Quantity62

Reason for Recall

A post market safety review following several customer complaints indicated additional clarification was necessary in the Midfoot Fusion Bolt Technique Guide to reduce potential for bolt back-out or migration by emphasizing the need for supplemental fixation.

Firm Action

An "Urgent Notice: Medical Device Recall" letter dated March 8, 2013 was sent to customers asking them to check their inventory, remove all affected product technical guides from use, and return them to Synthes along with the Verification form. Customer questions were directed to (610) 719-5450.

Root Cause

Labeling design

Distribution Pattern

Nationwide Distribution, including the states of MI, PA, SC, IA, MA, CA, WI, IL, and KS.