Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
FDA Recall Z-0014-2018 · Posted None
Recalling FirmHandicare AB
Recall StatusTerminated
Product CodeFSA
Date Initiated2016-07-28
Date Terminated2018-12-17
Product Quantity9 units
Reason for Recall
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin
smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause
increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
Firm Action
The firm initiated their recall by email on 07/28/2017. The firm directed the consignee to repair the units on site and destroy the affected parts.
Root Cause
Process control
Distribution Pattern
OH, PA, MD