Joerns Ultra Care 770 Model AC Powered Beds with model numbers U770, U770 AL , U770 GNDAL. Sunrise Medical, Long Term Care Division, 5001 Joerns Drive, Stevens Point, WI 54481 USA.
FDA Recall Z-0016-2007 · Posted 2006-10-05
Advertisement
Recalling FirmSunrise Medical CCG, Inc.
Recall StatusTerminated
Product CodeFNL
Date Initiated2006-01-01
Date Terminated2007-04-09
Product Quantity6,921
Reason for Recall
The control box used in the actuator systems appears to have a potential to become stuck in one position resulting a non-functioning bed.
Firm Action
The objective is to replace all of these systems. Phone contact of affected facilities began in January 2006. All facilities to be contacted by June 23, 2006. All products will be corrected at the location of the end-user by June 30, 2006.
Root Cause
Other
Distribution Pattern
Nationwide and Canada.