Siemens RAD Fluoro Uro Systems - Luminos dRF, Ysio and Uroskop Omnia with software versions VB10D, VB10F, VB10C and VB10E. Fluroscopic X-Ray Systems
FDA Recall Z-0016-2014 · Posted 2013-11-01
Advertisement
Recalling FirmSiemens Medical Solutions USA, Inc
Recall StatusTerminated
Product CodeOWB
Date Initiated2013-05-06
Date Terminated2014-11-06
Product Quantity23
Reason for Recall
Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition. Additionally, it is not possible for the user to stop the radiation manually; ho
Firm Action
Siemens initiated this correction on May 6, 2013 by sending Update InstructionXP005/13/S. Firm subsequently issued a letter on June 3, 2013 to all accounts.
1. Siemens Field Modification of the Luminos dRF, Ysio & Uroskop Omina running software versions VB10D, VB10F, VB10C and VB10E, which improves the signal transmission for the automatic exposure control, is available as of May 2013, and systems will be updated within one year.
2. The revised software fix for models Luminos dRF, Ysio & Uroskop Omina running software versions VB10D, VB10F, VB10C and VB10E, will be provided to all identified customers at no charge.
CDRH approves the CAP subject to the following conditions:
1. Notification of all dealers and purchasers is to be made within 15 working
days of receipt of this letter in the manner specified in 21 CFR 1003.21 and
1003.22. This office and the Food and Drug Administration (FDA) district
office coordinator noted below are to be included in the notification.
2. Siemens Medical Solutions USA, Inc., will notify the FDA district office in
Philadelphia, Pennsylvania, with reports of progress regarding this field
correction.
Root Cause
Software design
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.