Vista Dental Products Chlor-XTRA - 12mL Prefilled Syringes, QTY: 8, REF 503850
FDA Recall Z-0017-2020 · Posted 2019-10-01
Advertisement
Recalling FirmInter-Med Llc
Recall StatusTerminated
Product CodeKJJ
Date Initiated2019-06-27
Date Terminated2021-08-10
Product Quantity384 units
Reason for Recall
The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
Firm Action
The firm initiated the recall by letter on 06/27/2019. The letter directed the consignee to quarantine the product and contact the returns department for return instructions.
Root Cause
Process control
Distribution Pattern
US, South Africa, Switzerland, Poland, Italy, Singapore, Australia
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.