Oxylog 3000, Emergency and Transport Ventilator; Catalog #2M86300; Drager Medical Inc., 3135 Quarry Road, Telford, PA 18969.
FDA Recall Z-0018-2011 · Posted 2010-10-07
Advertisement
Recalling FirmDraeger Medical Systems, Inc.
Recall StatusTerminated
Product CodeCBK
Date Initiated2010-08-19
Date Terminated2010-11-18
Product Quantity78 potentially afftected units
Reason for Recall
Patients received insufficient ventilation. The Oxylog 3000 Instructions are inadequate relating to dead space volume.
Firm Action
The firm sent Important Safety Notice Letter, dated August 2010, and Instructions for Use supplement to consignees. The letter explained the issue with the affected product. The letter asks customers to attach the Instructions for Use amendment sheet supplied with the letter to their corresponding Instructions for Use. The firm also states that a they will provide a dedicated pediatric breathing circuit for the Oxylog 3000. Questions regarding the letter should be directed to Michael Kelhart at 1-800-543-5047. Questions regarding the use of the device should be directed to DragerService Technical Support at 1-800-543-5047 or your local Drager Sales Representative.
Root Cause
Error in labeling
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.