SOMATOM Definition Edge; intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

FDA Recall Z-0019-2016 · Posted 2015-10-05

Recall StatusTerminated
Product CodeJAK
Date Initiated2015-08-14
Date Terminated2016-12-05
Product Quantity113 total

Reason for Recall

software bug issues for SW-Version VA48A_SP0. The following safety issues were resolved: 1) Correction to improve visual warning and error indication son the gantry display. 2) Correction to improve acquisition data in order to optimize image quality. 3) Correction to improve robustness and general system behavior in some exception handling procedures. 4) Correction to improve auto post processin

Firm Action

A customer advisory notice, dated 8/14/15, was sent to direct accounts advising them of the issue to observe the safety notice and comply with the corresponding measure until the update has been fully completed.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution.