Stryker Navigation System - Hip Module Patient Tracker, green; Model 6007-005-000.
FDA Recall Z-0020-05 · Posted 2004-10-19
Advertisement
Recalling FirmStryker Instruments, Instruments Div.
Recall StatusTerminated
Product CodeHAW
Date Initiated2004-07-14
Date Terminated2004-10-18
Reason for Recall
The tracker is not locking adequately to the anchoring pin, resulting in unacceptable movement, which will affected the patient tracker system.
Firm Action
Sales force was notified these pre-launch devices are not to be used for surgical use via e-mail and telephone calls on 7/14/04.
Root Cause
Other
Distribution Pattern
United States.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.