Terminated

Interis Allura BiPlane X-Ray System

FDA Recall Z-0022-06 · Posted 2005-08-31

Recall StatusTerminated
Product CodeJAK
Date Initiated2005-07-27
Date Terminated2008-05-20
Product Quantity43 units

Reason for Recall

1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.

Firm Action

A letter issued on 7/25/05 advising customers of problem. The letter advised that an upgrade will be performed.

Root Cause

Other

Distribution Pattern

Devices were distributed to hospitals/medical centers nationwide.