Interis Allura BiPlane X-Ray System
FDA Recall Z-0022-06 · Posted 2005-08-31
Advertisement
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeJAK
Date Initiated2005-07-27
Date Terminated2008-05-20
Product Quantity43 units
Reason for Recall
1) During positioning of the Lateral Stand the rotation and/or skew movement may travel unexpectedly in opposite direction; 2) movement may be faster than normal speed; 3) the stand may not stop immediately upon release of the joystick.
Firm Action
A letter issued on 7/25/05 advising customers of problem. The letter advised that an upgrade will be performed.
Root Cause
Other
Distribution Pattern
Devices were distributed to hospitals/medical centers nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.