Philips Healthcare Integris Allura 15 -12 Fluoroscopy x ray system
FDA Recall Z-0022-2015 · Posted 2014-10-09
Advertisement
Recalling FirmPhilips Medical Systems, Inc.
Recall StatusTerminated
Product CodeOWB
Date Initiated2014-08-08
Date Terminated2014-12-03
Product Quantityone system
Reason for Recall
Philips had discovered through a field service testing where a device failed to comply with a Federal standard.
Firm Action
Philips Healthcare discovered through field service testing, that the device was in violation. Firm service rep. visited site on 7/01/14 and corrected unit.
The field service engineer immediately reprogrammed the generator for the X-ray tube installed in the system and re-adapted the x-ray tube to bring the system back into compliance.
At this time the Failure to comply has been corrected and the customer has been notified. This correction is to be free of charge to the customer.
For further questions please call (978) 687-1501.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US distribution to FL
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.