Babi.Plus 12.5 cm H2O Pressure Relief Manifold with Gas Supply Tube and Oxygen Sensor Port, REF 2691, Rx Only, UDI: 04710810100802
FDA Recall Z-0022-2020 · Posted 2019-12-20
Advertisement
Recalling FirmGalemed Corporation
Recall StatusTerminated
Product CodeBZD
Date Initiated2019-05-30
Date Terminated2024-05-20
Product Quantity14964
Reason for Recall
Reports of pressure relief manifold venting gas below the standard pressure.
Firm Action
On 05/30/2019, Medical Device Recall notices were sent to customers via email instructing them to do the following: Within 1 business day, acknowledge receipt of this recall notice. As soon as possible, within 5 business days, after this notification, notify your customers: a) That a product recall been initiated on the affected products. b) Indicate the nature of the recall. 1. Request that the affected lot codes be returned. 2. Customers who do not or cannot return affected product should be cautioned of the events which has led to this recall. 3. Within 5 business days of this notice, provide a list of all customers contacted with product code and lot information.
The recalling firm will provide information on processing of affected product. In addition, they will provide an estimated date when properly validated product will be available for shipment. Customers with additional questions were encouraged to contact the firm at [email protected].
Another recall notice, dated 09/19/19, was sent to customers and posted on the recalling firm's website (http://www.galemed.com/). Customers were asked to return reworked product.
Root Cause
Device Design
Distribution Pattern
US distribution to distributors in the states of: NC and NV.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.