Stryker brand Rugged EZ-Pro R4 Ambulance Cot, model 6092.
FDA Recall Z-0023-05 · Posted 2004-10-20
Advertisement
Recalling FirmStryker Medical
Recall StatusTerminated
Product CodeFPO
Date Initiated2004-08-17
Date Terminated2004-10-19
Product Quantity15
Reason for Recall
Locking pins may be installed backward, and cot may drop from highest to lowest position when the users are not prepared for the drop.
Firm Action
Service representatives visited each customer, inspected the cots, and reinstalled the lock pins, if required, beginning on August 10, 2004.
Root Cause
Other
Distribution Pattern
Alabama, California, Florida, Kansas, Montana, Oklahoma, Texas and Virginia.