Abbott AxSYM system FSH Master Calibrators (LN 7A60-30), for in vitro diagnostic use.

FDA Recall Z-0023-2007 · Posted 2006-10-12

Recall StatusTerminated
Product CodeJIS
Date Initiated2006-09-07
Date Terminated2011-09-09
Product Quantity797

Reason for Recall

Following a customer complaint, Abbott found an atypical stability profile for the lot of the calibrators listed in this recall. The investigation to date has shown that both controls and patient results have shifted upwards over time together.

Firm Action

The Device Recall Letter dated September 07, 2006, and Customer Reply Form was sent via Federal Express Priority mail to all customers who received the affected lots.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Puerto Rico, Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, Honduras, Trinidad-Tobago, St. Vincent, Bahamas, Panama, and Cayman Islands.