Biomedix SELEC-3 I.V. Administration Set. Individually sealed into a Tyvek and LDPE/PET pouch. Secondary packaging includes 48 pouched units per box/case. STERILE
FDA Recall Z-0023-2015 · Posted 2014-10-06
Advertisement
Recalling FirmBiomedix, Inc.
Recall StatusTerminated
Product CodeFPA
Date Initiated2014-08-14
Date Terminated2014-11-13
Product Quantity2,928
Reason for Recall
Customer contacted Biomedix of complaint and returned the failed set along with 86 units of unopened stock (96 units originally shipped). Biomedix performed extensive investigation and found insufficient application of glue. Recall was initiated due to the rate of the failure.
Firm Action
Biomedix sent an PRODUCT RECALL NOTIFICATION letter dated August 14, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer.
Please contact us with the quantity remaining in inventory or available to be returned to our facility. All units returned to Biomedix will be replaced with the same SELEC-3 item or a similar design at no charge.
Please call (1-800-627-2765 ) for further questions.
Root Cause
Process control
Distribution Pattern
US Distribution including the states of OH, NC, and TX
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.