Terminated

Quanta Technologies L.L.C. brand REORDER # LAPKIT 3. Contents 1 -55060 Suction/Irrigation Valve 2- 30012 Insufflation Tubing set 3- 5010 Anti Fog Solution with Sponge. Manufactured for Quanta Technologies, L.L.C., Tucson, AZ.

FDA Recall Z-0024-04 · Posted 2003-10-15

Recalling FirmVital Concepts, Inc.
Recall StatusTerminated
Product CodeGCY
Date Initiated2003-07-11
Date Terminated2004-04-12
Product Quantity160

Reason for Recall

The suction valve may become stuck in the 'suction on' position under vacuum.

Firm Action

A request for product return was sent to each customer beginning on July 11, 2003.

Root Cause

Other

Distribution Pattern

Arizona, Illinois, Indiana, Michigan, Ohio, Pennsylvania, South Carolina, Texas, Utah, Washington, Canada, Israel and Japan