Zimmer brand Bigliani/Flatow Total Shoulder trabecular metal glenoid component, 52 mm dia. articular surface, med, sterile, for cemented use only in the USA, Cat. no. 00430604652.

FDA Recall Z-0027-06 · Posted 2005-10-07

Recalling FirmZimmer Inc.
Recall StatusTerminated
Product CodeKWS
Date Initiated2005-08-23
Date Terminated2005-12-08

Reason for Recall

The implant may fracture at the junction of the base and the trabecular metal keel if adequate support for the base has not been achieved by the host bone and/or bone cement.

Firm Action

Consignees were notified via letter dated August 23, 2005 or by email.

Root Cause

Other

Distribution Pattern

Nationwide, Australia, Canada, England, Germany and Italy.