Lumenis brand DuoTome SideLite 550 Micron Delivery System; laser systems for ablating soft tissue. Catalog Number: 0641-800-01; Manufactured by Lumenis, Inc., 2400 Condensa Street, Santa Clara, CA

FDA Recall Z-0028-2007 · Posted 2006-10-12

Recalling FirmLumenis Inc.
Recall StatusTerminated
Product CodeGEX
Date Initiated2006-08-29
Date Terminated2007-01-22
Product Quantity10,815

Reason for Recall

Device lacks the black indicator markings on the metal tip and there is the potential for fiber degradation (detachment of the metal cap, and the fiber lasing straight) during use.

Firm Action

On 8/26/06, the firm initiated the recall and its notification was via letters and phone calls requesting return of affected product.

Root Cause

Other

Distribution Pattern

Worldwide, including USA, Japan, Europe, Hong Kong, Russia and Canada.