BioFlo PICC with ENDEXO Technology Maximal Barrier Nursing Kit, 4F-55 cm, UPN Product No. H965750121, REF/Catalog No. 75-012, STERILE --- Intended for short or long term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood, for central venous pressure monitoring and for power injection of contrast media.

FDA Recall Z-0028-2015 · Posted 2014-10-07

Recalling FirmNavilyst Medical, Inc
Recall StatusTerminated
Product CodeLJS
Date Initiated2014-08-25
Date Terminated2017-03-10
Product Quantity125 units/pouches

Reason for Recall

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Firm Action

BD Medical sent an Urgent Voluntary Medical Device Recall letter dated July 22, 2014, to all affected customers via Federal Express on August 25, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to segregate and quarantine all affected product. Immediately discontinue shipment of the affected product. Complete the Recall Response Card form and fax it back to BD at 1-201-847-6990. If the affected product was further distributed cusotmers should identify their cusotmers and notify them immediately of the product recall. Customers with questions were instructed to contact BD Customer Service at 1-888-237-2762, option 3. For questions regarding this recall call 518-742-4571.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution