Nox T3s Recorder - Product Usage: intended for ambulatory recording of physiological signals during sleep.

FDA Recall Z-0031-2022 · Posted None

Recalling FirmNOX MEDICAL
Recall StatusTerminated
Product CodeMNR
Date Initiated2021-09-07
Date Terminated2023-05-01
Product Quantity128 units

Reason for Recall

Error in the device firmware results in the inaccurate detection of device position.

Firm Action

Written notification letter titled, "URGENT: Medical Device Recall" was sent to all consignees on 09/07/2021. The letter instructs the consignee to follow the following instructions: "Further distribution or use of the affected devices should cease immediately. Devices should be quarantined pending device firmware update, per the instructions below. Due to the potential of inaccurate position-related reporting of TST, TRT, AHI and REI; studies performed with the affected T3s devices should be reanalyzed to determine if this has clinical impact on the patient diagnoses. A Nox Medical technical representative will contact you within 5 business days to schedule a remote session (via videoconference or other means) to upgrade the device firmware. Nox Medical requests that you formally acknowledge receipt of this notice and return the attached Customer / Distributor - Acknowledgement and Receipt Form (Attachment 1). This recall notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. If any affected devices which were in your possession (indicated in Attachment 2: Product Distribution and Information Table) have been transferred, please indicate this and provide as much detail as possible in the Acknowledgement and Receipt Form (Attachment 1)."

Root Cause

Finished device change control

Distribution Pattern

US Nationwide distribution in the states of OH, IL, GA, IA, CA, PA, NM, MI, SD.