Gambro Prismaflex Hemodialysis Machine, Catalogue Number 6023014700, Gambro Dasco S.p.A, Italy.

FDA Recall Z-0032-2007 · Posted 2006-10-12

Recall StatusTerminated
Product CodeKDI
Date Initiated2006-05-16
Date Terminated2012-01-03
Product Quantity153 units (US)

Reason for Recall

A potential risk of infusing air or infusion fluid to the patient during dialysis. Also, a possibility of an incorrect scale reading of the amount of patient fluid removal.

Firm Action

Consignees were notified by letter on 05/16/2006.

Root Cause

Software design

Distribution Pattern

Nationwide.