MEDLINE AUTOMATIC DIGITAL BLOOD PRESSURE UNIT, REF MDS1001

FDA Recall Z-0032-2025 · Posted 2024-10-08

Recall StatusOpen, Classified
Product CodeDXN
Date Initiated2024-08-30
Product Quantity52,521 units

Reason for Recall

It has been determined that certain Blood Pressure Monitors containing a faulty microchip have failed to power on or have shown display screen issues after the initial few uses, as well as overheating at times making the unit warm to the touch.

Firm Action

Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 08/30/2024 via email and USPS first class mail. The notice explained the problem, risk to user and requested that units demonstrating the failure be discarded. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. Medline issued a follow up notification to distributors on 09/13/2024. The notification contained the same information as the initial letter, but also instructed distributors to notify customers if the product was further distributed.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Panama and Jamaica.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.