2008T HD SYS. CDX W/BIBAG BLUESTAR
FDA Recall Z-0033-2024 · Posted 2023-10-06
Advertisement
Recalling FirmFresenius Medical Care Holdings, Inc.
Recall StatusOpen, Classified
Product CodeKDI
Date Initiated2023-09-06
Product Quantity733 units
Reason for Recall
Potential PCBA leaching from tubing of hemodialysis machines.
Firm Action
Consignees were sent an URGENT MEDICAL DEVICE CORRECTION NOTICE via USPS, dated 09/06/2023. The notice informed consignees that there is a potential for NDL-PCBAs could leach from silicone tubing manufactured with a peroxide-based catalyst. The firm also states that it has been demonstrated that dialysate run through 2008 Series machines used during typical maintenance HD treatments for at least 36 days, cumulatively, showed levels of NDL-PCBAs that were below toxicologically relevant thresholds. Consignees are to contact FMCRTG Technical Service at 1-800-227-2572 to schedule an update to devices to receive the correct tubing or if they have any further questions. Consignees with clinical questions are to contact Global Medical Information and Education Office at 1-855-2309. Consignees are to return the provided response form via fax or mail.
Root Cause
Device Design
Distribution Pattern
Domestic: Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.