Medtronic, Simplera Sensor, REF: MMT-5100JD1
FDA Recall Z-0033-2026 · Posted 2025-10-02
Advertisement
Recalling FirmMedtronic MiniMed, Inc.
Recall StatusOpen, Classified
Product CodeMDS
Date Initiated2025-05-07
Product Quantity1,440 units
Reason for Recall
The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
Firm Action
On May 7, 2025, Medtronic issued a "Urgent Field Safety Notice" to affected consignees. Medtronic asked consignees to take the following actions:
1. Verify if the sensors you currently have are from the impacted lot.
2. If you have a Simplera" sensor from the impacted lot, do not use this Simplera" sensor and dispose of it according to local regulations.
3. Please contact our helpline/your Medtronic contact at 01923 205167 to request a replacement.
4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter, even if your sensor is unaffected, or if you do not have any sensors from the affected lot.
Root Cause
Process design
Distribution Pattern
US: No distribution OUS: Austria, Belgium, Italy, Spain, Switzerland, United Kingdom
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.