The device's marketing brochure states: "IntelliVue M3150 Information Center...Philips Medical Systems". The beige-colored PAS-210 external speakers can be identified by their use of DB15-to-3.5 mm mini-phone jack cable (Agispec-081Y-4 or Agispec-081-6) to connect to the rp5700 PC. Indicated for central monitoring of multiple adult, pediatric, and neonatal patients.

FDA Recall Z-0034-2011 · Posted 2010-10-08

Recall StatusTerminated
Product CodeDSI
Date Initiated2008-11-21
Date Terminated2010-10-12
Product Quantity424 units in US.

Reason for Recall

Potential for Internal Sound Amplifier to malfunction which could delay treatment.

Firm Action

On 11/20/08, Philips began sending the URGENT-Medical Device Correction letter to their consignees via e-mails. On 11/21/08, the firm sent out the recall letter via United State Postal Service. The letter states the affected device as Philips IntelliVue Information Centers, running on the HP rp5700 PC platform equipped with an active speaker solutions using a DB-15 to 3.5 mm cable set. The consignees are advised to ensure that the audible alarm system is clear, free of noise and distortion and adjustable to different sound levels which is explained in p. 5-47 "Adjusting Alarm Tone Volume" of IntelliVue of the Instruction for Use, Release K Manual, supplied with the hardware, Philips Part Number 453564062301. Consignees can call Philips Healthcare Customer Care Service Center at 1-800-722-9377, #3, #1 and reference FCO 99862012.

Root Cause

Component change control

Distribution Pattern

Nationwide Distribution

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.